Lucille
Spinning disk confocal on an inverted fluorescence microscope.
Previous Configurations
Imaging modalities
Hardware Components
This instrument uses Class 3b lasers. All users must attend laser safety training.
Filter Spectra
Filters from Chroma
Compare the fluorescence spectra of your fluorophore to the filters on this microscope.
Illumination Setting | Excitation | Dichroic | Emission |
---|---|---|---|
442 | 442nm | 447/515/642 Triple | 480/40 |
488 | 488nm | 405/491/561/642 Quad | 525/50 |
515 | 515nm | 447/515/642 Triple | 540/30 |
561 | 561nm | 405/491/561/642 Quad | 620/60 |
642 | 642nm | 405/491/561/642 Quad | 700/75 |
Widefield DAPI | 395/25 | 400lp | 460/25 |
Widefield FITC | 480/40 | 505lp | 535/50 |
Widefield TRITC | 560/40 | 595lp | 630/60 |